Model: YTUP-10
System water production (25 °): 10 liters/hour
System process: PP+UDF+CTO+RO+DI+UP
Water inlet requirements: Urban tap water TDS<250ppm, 5-45 ° C, 0.02-0.25Mpa, pH3-10
Pre processing unit: 10 inch PP filter element X1+10 inch granular activated carbon filter element X1+10 inch sintered activated carbon filter element X1
RO reverse osmosis unit: 75GPDRO reverse osmosis membrane X1
Post purification unit: Ultra purification column X4
Resistivity/conductivity: 18.25M Ω .cm@25 ℃/0.054um/ cm@25 ℃ (<0.1um/cm)
Heavy metal ions:<0.1ppb
Total Organic Carbon (TOC):<10ppb
Heat source/endotoxin:-
Microbial bacteria:<1CFU/ml
Ribonuclease:-
Deoxyribonuclease:-
Particulate matter (>0.22 µ m):<1/ml
TDS value: RO water TDS (total solid solubility ppm) < influent TDS × 4% (desalination rate>99%)
Separation rate of divalent ions: 95% -99% (when using a new RO membrane)
Organic matter separation rate:>99%, when MW>200Dalton
Particle and bacterial separation rate:>99%
Maximum production capacity of reverse osmosis water: 2.0 liters/minute (opening pressure water storage bucket)
Front water outlet: 2; RO reverse osmosis nozzle, UP ultrapure nozzle
Side outlet: water inlet, pressure bucket, wastewater outlet
Working pressure monitoring: 0-1Mpa, 0-150PSI
Digital water quality monitoring: TDS test pen+LCD online resistivity, conductivity, TDS, water temperature monitoring
Overall dimensions (length × wide × Height: 44X45X46cm
Power supply: 220V, 50Hz
Power: 30W
Weight: 25kg
Standard configuration: The host includes 1 set of purification column, TDS test pen, and accessory package (optional) 3.2G external water bucket
- Product Details
- Product manual
Product use
It is widely used for qualitative/quantitative analysis items in routine experiments such as laboratory ware washing, reagent preparation, microbiological examination, biochemical analysis, etc.
Product features
1.Full-automatic pressure sensor and microcomputer control work to realize automatic production of pure water, humanized operation display system, with audible and visual alarm function;
2. The whole pipeline adopts the quick-plug interface, the standard is equipped with the water supply port of the external equipment, and the external bucket can be added, and the storage bucket of various specifications can meet different needs;
3. All pipelines are certified by NSF and adopt modular and quick-connect design, which is convenient for installation and maintenance;
4. The water quality of the water source is low, and the carefully designed pretreatment system can effectively treat different raw water in different places;
5. High water yield, long service life of consumables and low operating cost;
6. No water alarm, water full alarm, ultra-pure water (parameters can be set at will) over-limit alarm;
7. Full-automatic RO membrane anti-scaling washing procedure to extend the service life of RO membrane;
8. Water and electricity separation, overcurrent protection device, water inlet protection explosion-proof device and other protective devices make it safer to use water and electricity;
9. Portable TDS test pen, dry cell design, can measure TDS total solid solubility (approximate conductivity) of tap water and RO water at any time;
10. High brightness backlit LCD online resistivity/conductivity/TDS value/water temperature monitor, with an accuracy of 0.01M Ω. cm, to monitor the water quality in real time;
11. Electronic mixed bed resin, large capacity purification tank design, always ensure the stability of top water quality and water use;
12. (0.45+0.1) um medical-grade terminal bag type sterilization filter ensures absolute sterility of water quality;