Product parameters:
Model: YT-1096A
Light source: DC12V 22W; Imported halogen tungsten lamp
Optical path system: 8-channel vertical optical path system
Wavelength range: 400-1000nm
Filter: Standard 405nm, 450nm, 492nm, 630nm, other wavelengths optional, can load 10 filters
Reading range: 0-4.000Abs
Resolution: 0.001Abs
Indication error: ≤+0.01Abs
Stability: ≤+0.002Abs
Repeatability: ≤ 0.2%
Vibration plate function: Level 3 vibration plate speed and time adjustable from 0 to 255 seconds
Display operation: 8-hour color LCD screen, visible speaker board, display of whole board information, touch screen operation
Workstation: A professional enzyme-linked immunosorbent assay software that can store 200 sets of programs and provide rich calculation modes such as absorbance, threshold, linear equation, logarithmic equation, quadratic equation, cubic equation, exponential curve, power function, absorbance percentage concentration logarithmic equation, 4-parameter equation, etc
Power supply: Wide voltage design AC100-240V 50-60HZ
Net weight: approximately 11Kg
Size: 433 (L) * 320 (w) * 308 (H) mm
Injection: Automatic opening/closing of the injection chamber door, no need for manual opening of the lid
Material: The bottom plate of the enzyme-linked immunosorbent assay instrument is made of stainless steel material
- Product Details
- Product manual
Ye Tuo YT-10096A ELISA reader
Model: YT-10096A
Product Overview:
A widely used experimental technique for detecting biomolecules. It is based on the principle of enzymatic biochemical reactions and is commonly used to determine the concentration of biological molecules such as proteins, antibodies, stressors, antigens, etc. in samples.
Product characteristics of Ye Tuo YT-10096A enzyme-linked immunosorbent assay analyzer
Industrial grade color LCD display, touch screen operation
Eight channel fiber optic measurement system, imported detector
Accurate center positioning function, accurate and reliable
Three types of linear vibration plate functions, adjustable speed and time
● Cut Off method to determine the formula and input what you have entered
● End point method, two-point method, dynamics, single/dual wavelength testing mode